CR 4200

Overview

5

ba76121e01

09/2012

1

Overview

The thermoreactor CR 4200 is a dry temperature control device for
laboratory use. It facilitates and secures the digestion using reaction
cells.

The thermoreactor has 8 fixed temperature programs.

 1: 148 °C for 120 minutes
 2: 120 °C for 30 minutes
 3: 120 °C for 60 minutes
 4: 120 °C for 120 minutes
 5: 100 °C for 60 minutes
 6: 148 °C for 20 minutes
 7: 150 °C for 120 minutes
 8: 100 °C for 30 minutes

8 more temperature programs can be set up freely. The reaction
temperature can be adjusted from room temperature to 170 °C, the
heating time from 0 to 180 min.

The thermoreactor takes 24 reaction cells with an outer diameter of
16 mm.

The thermoreactor CR 4200 has two independent thermoblocks and
thus enables two temperature programs to run simultaneously.