8 what to do if, 1 information on errors, 2 error causes and remedies – YSI IQ S ENSOR N ET System 182 User Manual

Page 127: What to do if -1, Information on errors -1, Error causes and remedies -1, Chapter 8 w, 8what to do if

Advertising
background image

System 182

What to do if ...

8 - 1

ba76022e01

12/2011

8

What to do if ...

8.1

Information on errors

Log book

The IQ S

ENSOR

N

ET

system performs a comprehensive cyclical self

test during operation. While doing so, the system identifies all states
that deviate from normal operation and enters corresponding
messages in the log book (information or error message).

With the aid of the log book, you can call up instructions on how to clear
the error directly on the terminal. The log book is described in detail in
section 4.4.3 S

ENSOR

STATUS

S

XX

.

Note
Information on possible errors in IQ sensors and MIQ output modules
is given in the chapter W

HAT

TO

DO

IF

... of the respective component

operating manual.

8.2

Error causes and remedies

The system does no

longer react on entries

"Error" in the measured

value display

The selected system

language was not

activated for a sensor

Cause

Remedy

– System error

Reset the system:
– Switch off the power supply

and switch it on again after
10 s

Cause

Remedy

– Communication with the

IQ Sensor interrupted

– Check the cable connection

– Error in the IQ sensor

– Disconnect the IQ sensor

and reconnect it after 10 s

Cause

Remedy

– A system language was selected

that is not available in the
sensor.
The English language was
activated instead of the selected
system language.

– Contact YSI,

a software update is
required for the respective
components

Advertising