Commissioning

AmmoLyt

®Plus

700 IQ

3 - 10

ba76003e01

01/2012

3.5.2

Setting table of the AmmoLyt+

Menu item

Selection/values

Explanations

Measuring mode

NH4-N

NH4

mV

Citation form of the mass concentration or
voltage of the electrode.

Measuring range

(Measuring mode: NH4-N)

AutoRange

0.1 ... 100.0 mg/l

1 ... 1000 mg/l

2 measuring ranges can be selected.
With AutoRange, the instrument
automatically switches to the suitable
measuring range.

Measuring range

(Measuring mode: NH4)

AutoRange

0.1 ... 129.0 mg/l

1 ... 1290 mg/l

2 measuring ranges can be selected.
With AutoRange, the instrument
automatically switches to the suitable
measuring range.

Measuring range

(Measuring mode: mV)

-2000 ... 2000 mV

Fixed range

AmmoLyt+K

hide

display

hide (standard setting):

The potassium electrode is not displayed
as an extra sensor in the measured value
display.

display:
If the sensor is equipped with the
potassium electrode, the electrode is
displayed as the extra sensor AmmoLyt+K
in the measured value display.

The relevant settings can be done in an
extra setting table (see section 3.5.3).
Note: If the maximum number of sensors
for the system is exceeded by activating
the sensor, the AmmoLyt+K cannot be
activated.

Calib. history K
(only with the setting
AmmoLyt+K: hide)

Do not download

Transmit to log book

Transmit to log book generates a log book
message with the calibration history of the
potassium electrode. When opening the
setting table again the setting is reset to
Do not download.