Setting table of the am – YSI AmmoLyt Plus 700 IQ User Manual

Page 25

Advertising
background image

AmmoLyt

®Plus

700 IQ

Commissioning

3 - 13

ba76003e01

01/2012

3.5.3

Setting table of the AmmoLyt+K (potassium sensor)

Note
These sensor settings are only available if in the sensor settings for the
AmmoLyt+ the AmmoLyt+K setting was set to active (see section
3.5.2).

The basic settings are taken over from the AmmoLyt+ sensor but can
be adjusted separately afterwards.

Menu item

Selection/values

Explanations

Measuring mode

K (mg/l)

mV

Citation form of the mass concentration or
voltage of the electrode.

Measuring range

(Measuring mode: K (mg/l))

1 ... 1000 mg/l

Fixed range

Measuring range

(Measuring mode: mV)

-2000 ... 2000 mV

Fixed range

IQ-LabLink

not active

(not used)

Temperature mode
(only with

Measuring mode: K (mg/l))

°C

°F

Unit of the measured temperature value
(Celsius, Fahrenheit).

Temp. adjustment
(only with

Measuring mode: K (mg/l))

-1.5 °C ... +1.5 °C

The temperature compensation function
enables the temperature sensor to be
balanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 °C).
Notes:

Due to the thermal capacity of the
sensor, it is necessary to place it in a
container with at least 2 liters of water.

Leave the sensor in this container for at
least 15 minutes while stirring
occasionally, then carry out the
adjustment.

If the temperature difference of the
water and sensor is > 10°C, leave the
sensor in the container for at least one
hour while stirring occasionally.

Conc. offset

-10 mg/l ... +10 mg/l

Constant value, which is added to the
measuered value.