9 maintenance condition of sensors, Maintenance condition of sensors -41, Section 4.9) – YSI IQ SensorNet DIQ/CR3 User Manual

Page 55

Advertising
background image

MIQ/CR3; DIQ/CR3

Settings

4 - 41

ba76032e01

01/2012

4.9

Maintenance condition of sensors

The maintenance condition of sensors serves to calibrate, clean,
service and repair (remove and replace) sensors.

In the maintenance condition

the system does not react to the current measured value or the
condition of the selected sensor

linked outputs are frozen

sensor errors do not prompt changes in the conditions of linked
outputs.

The maintenance condition is automatically activated

during calibration. After calibration the sensor remains in the
maintenance condition until the maintenance condition is switched
off manually (see section 4.9.2)

during the cleaning cycle (see section 4.5.7)

Switch on the maintenance condition manually when you want to clean,
service or repair (remove and replace) a sensor (see section 4.9.1).

When you have finished cleaning, servicing or repairing the sensor,
switch off the maintenance condition manually (see section 4.9.2).

Note
When a sensor is in the maintenance condition, the measured values
or status indications of the sensor in the measured value display flash.