YSI IQ SensorNet DIQ/CR3 User Manual

Page 60

Advertising
background image

Technical data

MIQ/CR3; DIQ/CR3

6 - 2

ba76032e01

01/2012

Relays

Current outputs

Output

Physically separated from the
IQ S

ENSOR

N

ET

Max. switching voltage

250 VAC or 24 VDC

Max. switching current

5 A (AC and DC)

Installation
requirements

Fuse rating on the operator side: Maximum
5 A

Relay functions

System monitoring

Sensor monitoring

Limit indicator

Frequency controller

Pulse-width contr.

Cleaning

Sensor-controlled

Manual control

Alarm contact (MIQ/CR3 only)

Output

Physically separated from the
IQ S

ENSOR

N

ET

Output current

Can be switched between 0 - 20 mA and
4 - 20 mA
In the case of errors, can be set to: 0 ... 21
mA

Max. initial output
voltage

15 V, in the case of missing or incorrect
burden

Max. load

500

Ω

Accuracy

0.3 % ± 50

μ

A

Functions

Programmable as:

– Analog output

– PID controller

– Output with fixed current value (for test

purposes)