4 limit indicator, See section 4.5.4 – YSI IQ SensorNet DIQ/CR3 User Manual

Page 33

Advertising
background image

MIQ/CR3; DIQ/CR3

Settings

4 - 19

ba76032e01

01/2012

General

sensor errors

4.5.4

Limit indicator

Function

The characteristic of the limit indicator is laid down in the Limit value
UL
, Limit value LL, Hysteresis UL and Hysteresis LL settings. The
fundamentals of the function are described in the introductory chapter
(see section 4.1.2).

In order to set up the Limit indicator function for a relay output, the relay
output must be linked with a sensor (see section 4.3).

Settings

Init

can prompt a relay action for a short time, depending
on the starting behavior of the system

----

Invalid measured value, or defective sensor

Error

Communication with sensor interrupted

OFL

Measuring range undercut or exceeded (overflow)

Setting

Selection/Values

Explanation

Limit values

UL main variable
LL main variable

UL adjoining var.
LL adjoining var.

Main variable
designates the actual
measured parameter
of the sensor (e.g.
pH, oxygen, etc.).
Adjoining variable
designates an
additional measured
parameter (e.g.
temperature).

Limit value UL

Any upper and lower
limiting value
within the measuring
range (sensor-
dependent)

Minimum spacing
between the upper
and lower limiting
value:
5 % of the measuring
range

Limit value LL

Hysteresis UL

0 - 5% of the
measuring range

Hysteresis for Limit
value UL
and Limit
value LL
.

Hysteresis LL