Installation

3 - 3

ba76032e01

01/2012

General installation

instructions

Observe the following points when attaching connecting wires to the
terminal strip

Shorten all wires to be used to the length required for the installation

Always fit all the ends of the wires with wire end sleeves before
connecting them to the terminal strip

Any wires that are not used and project into the enclosure must be
cut off as closely as possible to the cable gland.

Screw a small cable gland with sealing ring into each remaining free
opening and close it with a blind plug.

Warning

No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages which could result in life
threatening electric shock when working with the IQ S

ENSOR

N

ET

.

Always cut off any wires that are not in use as closely as possible
to the cable gland.