Membrane type – YSI Pro20 User Manual

Page 17

Advertising
background image

MEMBRANE TYPE

Membrane Type sets the type of membrane used on the dissolved oxygen sensor;
either 1.25 PE (Yellow) or 2.0 PE (blue). Use the up or down arrow keys to

highlight Membrane Type and press enter

to open the membrane submenu.

Highlight the membrane type corresponding to the membrane installed on the
sensor and press enter to confirm. The enabled membrane type will have an ‘X’ in
the box next to it. Use the down arrow key to highlight the ESC – Exit box and
press enter to save changes and to close the membrane submenu.

The dissolved oxygen sensor is supplied with membranes specific to the sensor
type ordered and are color coded as described in the following tables.

Galvanic Membrane Kits

Item Color

Material

Description

5913

Yellow 1.25 mil polyethylene Faster response time and less flow

dependence than traditional Teflon®
membranes

5914

Blue

2.0 mil polyethylene

Less flow dependence than 1.25 mil
but somewhat slower response

Polarographic Membrane Kits

Item Color

Material

Description

5908

Yellow 1.25 mil polyethylene Faster response time and less flow

dependence than traditional Teflon®
membranes

5909

Blue

2.0 mil polyethylene

Less flow dependence than 1.25 mil
but somewhat slower response

Selecting a Dissolved Oxygen Membrane

Membrane Type

Flow Dependence After 4 Minutes

Typical Response Time

to 95%

5913, 5908 - Yellow 25%

8 seconds

5914, 5909 – Blue

18%

17 seconds

13