YSI Pro20 User Manual

Page 38

Advertising
background image

34

Symptom Possible

Solution

Instrument readings are inaccurate.

1) Verify correct sensor/membrane type

selection in the System Setup menu.

2) Verify calibration, barometer reading,

and salinity settings are correct and
recalibrate.

3) Verify accurate temperature readings.

4) Sample temperature is over 45 ºC, the

temperature compensation range.

5) Probe may not have been in 100%

water saturated air during calibration
procedure. Moisten sponge in
calibration sleeve and recalibrate.

6) Replace membrane and electrolyte,

recalibrate.

7) Clean sensor electrodes.

8) Return system for service.

Dissolved Oxygen values display
Over or Undr on run screen.

1) Verify correct sensor/membrane type

selection in the System Setup menu.

2) If using a polarographic sensor, allow

instrument to warm up for 5 – 15
minutes before use.

3) Sample O

2

concentration is more than

50 mg/L or 500%, or less than –0.02
mg/L or -0.3%.

4) Verify barometer and salinity settings

are correct and recalibrate.

5) Verify accurate temperature readings.

6) Replace membrane and electrolyte.

Recalibrate.

7) Clean sensor electrodes.

8) Return system for service.