Temperature values display Over or
Undr on run screen.

1) Sample temperature is less than -5º C

or more than +55

o

C . Increase or

decrease the sample temperature to
bring within the allowable range.

2) Return system for service.

S P E C I F I C A T I O N S

Parameter Range Resolution

Accuracy

-5 to 55 ºC *

0.1 ºC

± 0.3 ºC

Temperature

23 to 113 °F

0.1 °F

± 0.6 ºF

0 to 200% air
saturation

1% or 0.1%, user
selectable

± 2% of the reading or
2% air saturation,
whichever is greater

200 to 500%
air saturation

1% or 0.1%, user
selectable

± 6% of the reading

0 to 20 mg/L

0.1 or 0.01 mg/L,
user selectable

±2% of the reading or
0.2 mg/L, whichever is
greater

Dissolved Oxygen

20 to 50 mg/L

0.1 or 0.01 mg/L,
user selectable

±6% of the reading

Barometer

400.0 to 999.9
mmHg**

0.1 mmHg

± 5 mmHg within 5
°C of calibration
temperature***

* Automatic Dissolved Oxygen Temperature Compensation Range is -5 to 45 ºC

**Available barometer units include: mmHg, inHg, mbars, psi, or KPa

***For operating temperatures below 10 °C or above 40 °C, the barometer must be
recalibrated to maintain accuracy specification.