15

auto shutoff

Auto Shutoff powers the instrument off after a user specified time period.

Highlight Auto Shutoff and press enter. Using the numeric entry screen, enter

a value between 0 and 360 minutes. To disable auto shutoff, set the value to 0

(zero).

BacKlIGht

Backlight can be set to Automatic or Manual.

Automatic turns the backlight on when you

turn the instrument on and when you press any

key. Manual allows you to turn the backlight
on or off with the backlight key

. When in

Automatic mode, the instrument will turn the

backlight off after 60 seconds without any keys

being pressed. If a key is pressed during that

time, the instrument will “reset” the 60 second

time period. The lighted keypad will turn off

after approximately 20 seconds.

sW VersIon (softWare VersIon)

SW Version shows the instrument’s software version.

serIal #

Serial # shows the instrument’s serial number and allows you to match it with

the number engraved on the back of the instrument’s case. The Serial # is also

the default Unit ID.

unIt Id

Unit ID is used to identify instruments in the Data Manager software included

with your instrument. It is also used to identify GLP files, Site Lists, Configuration

Files, and Data files transferred from the instrument to the PC. The default Unit

ID is the Instrument’s serial number. To modify the Unit ID, highlight Unit ID,

press enter and use the alpha/numeric entry screen. The character limit is 16

characters.

O P T I C A L D I S S O LV E D O X Y G E N

( O D O ™ )

Each ProODO cable/probe assembly and replacement ODO

sensor cap includes an instruction sheet with important

information unique and specific to each individual sensing

cap. This instruction sheet is important because it includes

the calibration coefficients for the sensor cap. After using this

sheet for general probe setup, be sure to store it in a safe place

in case you need to reload these calibration coefficients in the

unlikely event that they are ever deleted from the probe.

See Initial Setup | Unpacking the ODO Probe for initial setup instructions.

It is important to always keep your sensor in a moist

environment so the sensor cap does not dry out. (See

Maintenance and Storage for more information.)

odo proBe setup

Press Probe

, Highlight DO, press enter.

Enabled allows you to enable or disable the DO

sensor. Highlight Enabled and press enter to

activate or deactivate dissolved oxygen.

The DO Setup menu also displays the ODO

probe’s SW Version and Serial #. This

information is programmed into the probe at

the factory and digitally sent to the instrument

when the cable is connected.

Local DO allows for localized DO%

measurements. This sets the calibration value

to 100% regardless of the altitude or barometric

pressure. Highlight Local DO and press enter

to enable or disable this function. Local DO

is a method for the ProODO to factor in the

barometric pressure on each DO measurement.

In essence, even if the barometric pressure changes you wouldn’t notice the

difference with the DO% readings in air-saturated water or water-saturated air.

i

i