Initial setup – YSI ProODO User Manual

Page 7

Advertising
background image

4

5

InItIal setup

Throughout the manual, the term “probe” refers to the end of the cable where

the sensor is located, the term “sensor” refers to the Optical Dissolved Oxygen

sensing portion of the cable/probe assembly, and the term “sensor cap” refers to

the removable sensing cap that is replaced about once per year (Figure 3).

Figure 3

UNPACKING THE ODO PROBE

Each ProODO cable/probe assembly and replacement ODO

sensor cap includes an instruction sheet with important

information unique and specific to each individual sensing cap.

Remove the cable/probe assembly from the shipping container and locate

the instruction sheet included with your assembly. This instruction sheet is

important because it includes the calibration coefficients for your sensor cap.

After using this sheet for general probe setup, be sure to store it in a safe place in

case you need to reload these calibration coefficients in the unlikely event that

they are ever deleted from the probe.

Note - A new cable/probe assembly already has a

sensor cap installed and the sensor cap coefficients

are preloaded into the probe at the factory.

Preparing the probe for the first time:

Remove the metal probe guard from the probe by turning it

1.

counterclockwise.

Remove the red storage cap which contains a moist sponge from the end of

2.

the probe by pulling it straight off the sensor. Save this to use later for long

term storage.

Sensor

Probe

Sensor Cap

i

i

Reinstall the probe guard by sliding it carefully over the sensor and then

3.

threading it onto the cable/probe assembly with a clockwise rotation (Figure

4).

Locate the grey calibration/storage sleeve that was shipped with your probe/

4.

cable assembly. Moisten the sponge in the grey calibration/storage with a

small amount of clean water.

Slide the calibration/storage sleeve over the probe guard to keep the probe

5.

in a moist atmosphere for storage or calibration (Figure 5). It is important

to always keep your sensor in a moist environment so the sensor cap does

not dry out. (See Care, Maintenance, and Storage for more information.)

Figure 4

Figure 5

CONNECTING THE PROBE/CABLE ASSEMBLY TO THE

INSTRUMENT

To connect the cable, align the keys on the cable connector to the slots on the

instrument connector. Push together firmly, then twist the outer ring until it

locks into place (Figure 6). This MS-8 (Military Spec) connection is water-

proof.

Figure 6. Note the keyed connector. The cable and instrument

connectors can only be mated once the keyed sections are properly

aligned. When disconnected, the cable connector and the

connector on the instrument maintain an IP-67 rating.

Advertising