Laboratory barometer readings are usually “true”

(uncorrected) values of air pressure and can be used “as is”

for barometer calibration. Weather service readings are

usually not “true”, i.e., they are corrected to sea level, and

therefore cannot be used until they are “uncorrected”. An

approximate formula for this “uncorrection” is below:

True BP = [Corrected BP] – [2.5 * (Local Altitude in ft. above sea level/100)]

T E M P E R A T U R E

All probe/cable assemblies have built-in temperature. Temperature calibration
is not required nor is it available. To set the units, press Probe

, and select

Display. Highlight Temperature and press enter. Highlight the desired

temperature units of °F, °C, or K and press enter to confirm the selection. Only

one temperature unit may be displayed at a time. You may also choose to not

display temperature. If you choose not to display temperature, dissolved oxygen

readings will still be temperature compensated.

T A K I N G M E A S U R E M E N T S

To take readings, insert the probe into the sample. Move the probe in the

1.

sample to release any air bubbles and to provide a fresh sample to the sensor

cap. This movement is only necessary when first inserting the probe into

the sample. Since the ProODO utilizes optical luminescent technology,

continuous sample movement or stirring is not required. The probe will

fit into a 300 mL BOD bottle for taking initial and final BOD readings. For

best results in a BOD bottle, a stirring device should be used to properly mix

the sample and to keep solids from settling at the bottom.

Allow the temperature readings to stabilize and wait approximately 25-35

2.

seconds for the DO readings to stabilize.

Log One Sample is already highlighted in Run mode. Press Enter to open

3.

a submenu. Highlight Sites or Folders and press Enter to select the site or

folder to log the sample to.

If necessary, use the keypad to create a new Site or Folder name. If Site List

4.

and Folder List are disabled in the System menu, you will not see these

options when logging a sample.

i

Once the Site and/or Folder name is selected, highlight

5.

Log Now and press

Enter. The instrument will confirm that the data point was successfully

logged.

If you would like to log at a specific interval vs. logging one sample at a time
or vice versa, press System

, then highlight Logging and press Enter. Select

Continuous Mode and adjust the time Interval if necessary. On the Run screen,

the option to log will change from Log One Sample to Start Logging based on the

time interval entered in the Logging Menu. Note – When utilizing the Manual

sampling mode, the Continuous Logging Interval must be set to 10 seconds or

greater.

During a continuous log, the Start Logging dialog box on the Run screen will

change to Stop Logging. Press Enter to stop continuous logging.

Note - There is NO WARM-UP period associated with the ProODO sensor so

you may wish to turn off the ProODO instrument between readings to conserve

battery power.

F I L E S A N D S I T E L I S T S

fIle MeMory

To view the file memory, press File

.

The Data Memory shows a percentage

indicating the amount of memory available.

If the file memory is near 0%, files should be

downloaded to a PC and/or deleted to free up

memory.

VIeW data

Press File

, highlight View Data, press enter.

Configuring your data view:

Site: will allow you to view data from one

particular site or all sites. Highlight Site, press

enter, and select the site you wish to view data

from or select All Sites to view data from all

sites.