Restore default calibration values, Accessories / part numbers – YSI ProODO User Manual

Page 23

Advertising
background image

36

37

connection issue between the cable and instrument. The following are some of

the potential messages:

Message

Description and Recommended Action

Probe Temp over range

Temperature is over

100° C or reading

erroneously. Check cable connection and

ensure Temp sensor is clean.

Probe Temp under range

Temperature is under

-10° C or reading

erroneously. Check cable connection and

ensure Temp sensor is clean.

DO over range

DO% saturation is over 550%. Check sensor

cap and recalibrate.

DO under range

DO% saturation is under -5%. Check sensor

cap and recalibrate.

Barometer over range

Barometric pressure is reading over 988

mmHg. Calibrate barometer.

Barometer under range

Barometric pressure is reading under 375

mmHg. Calibrate barometer.

ODO Communications

Error

No communication between the instrument

and cable. Check cable connection.

Clock Battery Low

Internal battery for real time clock has low

voltage. Contact Technical Support.

Measurements Locked!

Measurements are held in Manual Sampling

mode. Select Update Measurements or Log

Held Data.

Illegal Value may appear during alpha/numeric entry on the message line. This

only appears if the values entered do not match the formatting. This will also

appear in the GLP security area if the password is incorrect.

restore default calIBratIon Values

Occasionally, the instrument may need to have the factory calibration default

values restored. In order to accomplish this press Calibrate

, highlight

Restore Default Cal and press enter. Select the parameter you wish to restore,

either DO or Barometer, and press enter. After selecting barometer or DO, you

will be asked to confirm the operation. Highlight Yes and press enter.

A C C E S S O R I E S / PA R T N U M B E R S

Part Number

Description

626281

ProODO Instrument

626250-1, -4, -10, -20, -30,

-40, -50, -60, or -100

1, 4, 10, 20, 30, 40, 50, 60, or

100-meter probe/cable assembly*

626320

Replacement Sensor Cap

605604

Communications saddle kit

605515

Data Manager desktop software

603075

Carrying case, soft-sided

603076 and 603078

Extended length flow cell and single port

adapter. Both items are required for use with

a ProODO