12

13

lanGuaGe

To change the instrument’s language setting,

highlight Language in the System menu and

press Enter. Highlight the desired language

and press enter to confirm. Languages include

English, Spanish, French, German, Italian,

Portuguese, and Norwegian.

radIx poInt

Radix Point allows the user to select between

a comma or a decimal in numeric radix point

for displayed values. For example, 1.00 becomes

1,00 when Use Comma is selected. Highlight

Use Decimal or Use Comma and press enter to

make your selection.

loGGInG

From the System menu, highlight Logging and press enter to view or change

the logging options. Logging options include Use Site List, Use Folder List,

Continuous Mode, and Interval.

Use Site List and Use Folder List are optional

ways of filing or ‘tagging’ your logged data

points. If these settings are enabled, you will

be prompted to select a Site and/or Folder to

‘tag’ to logged data points. See the File and Site

Lists section of this manual for information on

creating Site and Folder Lists.

Check the box for Continuous Mode if you

want to log samples continuously at a specific

time interval. To set the length of time between

logged samples, highlight Interval and press

Enter. Enter the interval as HH:MM:SS. This

interval will display at the top of the screen

when you select the Start Logging option in

run mode. If operating the unit in Manual

mode, see next section, the continuous logging

interval must be set to 10 seconds or greater in order to log data.

To log one sample at a time, uncheck Continuous Mode.

saMplInG Mode

The Sampling mode can be set to Automatic or Manual. In Automatic mode,

which is the default mode, the readings are

constantly updated. In Manual mode, the

readings are held or ‘locked’ on the display

once they reach the stability of the Auto Stable

setting in the Probe menu. The user must then

press a key to either log the held data or update

the measurements. Selecting Manual mode may

increase the instrument’s battery life. Highlight

either Automatic or Manual and press enter to

make your selection. If continuously logging in

Manual mode, the logging interval must be set

to 10 seconds or greater.