Barometer, Display - barometer, Calibration - barometer – YSI ProODO User Manual

Page 16: Display - barometer calibration - barometer

Advertising
background image

22

23

an ID. When enabled these IDs are stored with each calibration record in the

GLP file.

After selecting the Probe ID and/or User ID if appropriate, highlight DO and

press enter. Highlight DO mg/L or ppm and press enter.

Highlight Calibration value and press enter to

manually input the sample’s dissolved oxygen

value. Highlight Accept Calibration and press

enter once the temperature and Dissolved

Oxygen readings have stabilized. If User Field 1

or 2 are enabled, you will be prompted to select
the fields and then press Cal

to complete the

calibration. After completing the calibration, the

message line will display “Calibration Successful”.

Press Esc

to cancel the calibration.

CALIBRATING IN MG/L OR PPM AS A TITRATION:

2-POINT CALIBRATION WITH ZERO SOLUTION

Place the sensor in a solution of zero DO. Be sure to immerse both the ODO and

Temperature sensor into the water.

A zero DO solution can be made by dissolving approximately 8 - 10 grams of

sodium sulfite (Na

2

SO

3

) into 500 mL tap water. Mix the solution thoroughly. It

may take the solution 60 minutes to be oxygen-free.

Press Calibration

. If Probe ID or User ID are enabled in the System GLP

menu you will be able to highlight these features and add, select, edit, or delete

an ID. When enabled these IDs are stored with each calibration record in the

GLP file.

After selecting the Probe ID and/or User ID if appropriate, highlight DO and

press enter. Highlight Zero and press enter. Wait for the temperature and

DO% values under “Actual Readings” to stabilize, then press enter to Accept

Calibration. If User Field 1 or 2 are enabled, you will be prompted to select the
fields and then Press Cal

to complete the calibration. The screen will then

prompt for a follow-up second point calibration.

Highlight the desired calibration units (mg/L or ppm) and press enter to

continue with the next point in a known titrated value. Rinse the sensor of

any zero oxygen solution using clean water. Follow the steps under One Point

Calibration for mg/L or ppm for the second point.

B A R O M E T E R

All ProODO instruments contain an internal barometer.

dIsplay - BaroMeter

Press Probe

, highlight Display, and press enter. Highlight Barometer, press

enter. The measurement unit options are: mmHg, inHg, mBar, PSI, kPa, or Atm.

Only one unit can be displayed at a time. Select None if you do not want to

display a barometric pressure reading.

Whether or not you choose to display the barometer reading, the barometric

pressure will still be used for calibrating DO and for compensating for pressure

changes if Local DO is enabled.

calIBratIon - BaroMeter

The barometer in the instrument is calibrated at the factory. If the barometer

requires calibration, press Calibrate

. If Probe ID or User ID are enabled in

the System GLP menu you will be able to highlight these features and add, select,

edit, or delete an ID. When enabled these IDs are stored with each calibration

record in the GLP file.

After selecting the Probe ID and/or User ID if appropriate, highlight barometer

and press enter.

Highlight the desired unit and press enter.

Highlight Calibration Value and press enter to

manually enter the correct “true” barometric

pressure. Next, highlight Accept Calibration,

and press enter. If User Field 1 or 2 are enabled,

you will be prompted to select the fields and then

press Cal

to complete the calibration. After

completing the calibration, the message line will

display “Calibration Successful”. Press Esc

to cancel the calibration.