Thank you for purchasing the YSI Professional Optical Dissolved Oxygen (Pro

ODO) instrument. The YSI ProODO

measures dissolved oxygen in water using

lifetime luminescence technology and uses a digital signal to send information

between the instrument and probe. Key advantages of the new ProODO include

the elimination of sensor flow dependence and sensor warm-up time, greater

stability, the ability to zero the sensor for more accurate measurements at low

dissolved oxygen levels, and the elimination of frequent membrane/electrolyte

changes. The ProODO also features a waterproof (IP-67) case, a rugged MS-8

cable connector, backlit display and keypad, user-selectable cable lengths, USB

connectivity, large memory with extensive site list capabilities, and a rugged,

rubber over-molded case. For product specification information, please visit

www.ysi.com or contact Technical Support at 800-897-4151 (+1 937 767-7241).

Reading the entire manual before use is recommended for an overall

understanding of the instrument’s features.

G E T T I N G S T A R T E D

InItIal InspectIon

Carefully unpack the instrument and accessories and inspect for damage.

Compare received parts with items on the packing list. If any parts or materials

are damaged or missing, contact YSI Customer Service at 800-897-4151 (+1 937

767-7241) or the authorized YSI distributor from whom the instrument was

purchased.

Battery InstallatIon

The ProODO uses 2 alkaline C-cell batteries. Battery life depends on sampling

mode and usage. When used in Manual Sampling mode, under normal

conditions, battery life is approximately 80 hours at room temperature. The

use of Automatic Sampling mode may slightly reduce the battery life. See the

System section of this manual for more information on Sampling Modes.

Figure 1. ProODO with battery cover removed. Notice battery symbols indicating

polarities.