Xylem FC220 OPEN CHANNEL FLOW MONITOR User Manual

Page 7

Advertising
background image

Global Water

800-876-1172

globalw.com

- 7 -

In these modes the user is either prompted to enter the pipe diameter or rectangular
channel width, in user programmable units. The WA and WP are automatically
calculated based on the diameter or width, and the measured water level.

The roughness factor is determined by the construction material, typical values are:

0.010

Plastic, other smooth surfaces

0.012

Smooth unpainted steel

0.013

Painted steel, coated cast iron

0.013

Clay drainage tile, smooth asphalt

0.013

Finished concrete

0.014

Vitrified clay sewer tile

0.014

Uncoated cast iron

0.017

Unfinished concrete

0.024

Corrugated metal storm drain


Polynomial:
A best fit 3

rd

order polynomial can also be used to calculate flow. This is useful when

you need to find an equation that best fits measured data or a lookup table. The
equation is in the form of:

Q = A + Bh + Ch² + Dh³

Where:
Q = Flow in cubic feet per second
A - D = User defined coefficients
h = Water level in feet

When selecting this mode, the user is prompted to enter the values for A, B, C and D.

Look-Up Table:
For applications where a single equation does not properly calculate flow, a single
look-up table can be programmed into the FC220. This is a factory programmable
option only. Data for this table must be provided to Global Water in two columns and
no more than 200 lines. The first column is water level in feet and the second column
is the flow in cubic feet per second at that water level. Liner interpolation is
automatically done when a measured water level is between two data points.