Communication protocol Serie30
Page 26/37
5.10
Function 74 : Read value of a channel (integer)
Request:
DevAddr 74
CH
CRC16_H CRC16_L
Response:
DevAddr 74
B3
B2
B1
B0
STAT
CRC16_H CRC16_L
Exception errors:
2
If CH > 5
3
If message length is incorrect
4
If a channel is in overflow/underflow/inactive state or the data is invalid
32
If device has not yet been initialised
Note:
Only in devices of Class.Group = 5.20.
Same as function 73, but values in 4-byte integer (long) B0 .. B3, where B3 is MSByte. The resolution is reduced to 0.1mbar.
Unit: CH0, P1 and P2: in Pascal (1Pa = 10
-5
bar).
T, TOB1 and TOB2: in 0.01°C
Status-Byte (STAT): See function 73.
For details in error-handling see chapter Error handling and recognition!