Alarm Times: Check the box

associated with each alarm to

enable that feature and enter

the desired start or stop time.

The logger will start at the next

occurrence of the Alarm 1 time and stop at the next occurrence of the Alarm 2 time

after the Alarm 1 timer has expired. Either or both alarms can be enabled

Note: Use care when programming the alarm times. If left enabled, changing

and reprogramming any of the setup parameters will restart the alarm timers

and could cause the logger to stop recording and the loss of important data.

XIV. Analog Setup:

The Analog Setup screen sets the engineering units and calibration numbers for each

of the analog channels, the endpoints of the bar graphs in the main screen and which

of the analog channels are enabled. These channels are calibrated in the same way

so a single channel will be discussed. It is important to note here that the information

stored in the data logger’s memory is just the sensors output level before it is

adjusted by the following calibration parameters. These parameters are used to

adjust the stored data for display and storage purposes only, at the time the data is

downloaded. If they are changed, the same stored data can be downloaded again

and the data will be scaled differently according to the new settings.