Xylem WL16 WATER LEVEL LOGGERS User Manual

Page 25

Advertising
background image

Global Water

800-876-1172 globalw.com

- 25 -

Disp High: This field sets the upper limit of the bar graph displayed in the main

screen and provides a way of scaling the graph beyond the upper limit used to

calibrate the logger. In this example, the upper calibration point was a water level of

8 feet but the maximum range of the sensor is 16 feet.

Disp Low: This field sets the lower limit of the bar graph displayed in the main

screen and provides a way of scaling the graph beyond the lower limit used to

calibrate the logger. In this example, the lower calibration point was a water level of

zero.

Decimal Places: From the pull-down menu, select the

number of decimal places to the right of the decimal

point that should appear in the historical data and in the

numerical data field at the bottom of the bar graph.

Calibrate: The “Calibrate” button provides a simple 4-step process for calibrating

the WL16 Data Logger to a particular sensor. Click the Calibrate button and follow

the menu instructions.

Click on the High EU field and enter the high

engineering unit that will be used to calibrate

the logger, not the highest reading the sensor

can output. In this example of water level, a

depth of 8 feet used to calibrate the logger,

even though the sensor can measure up to 15

feet.

Click Next.

Place the sensor in the condition that will

cause it to output the level corresponding

to the high EU entered in the previous menu.

Wait for the data reading to stabilize.

Click Next.

Advertising