Global Water

800-876-1172 globalw.com

- 26 -

Click on the Low EU field and enter the

low engineering unit that will be used

to calibrate the logger, not necessarily

the lowest reading the sensor can output.

In this example, the value is zero feet, which

is the lowest calibration point and water level.

Click Next.

Place the sensor in the condition that will

cause it to output the level corresponding

to the low EU entered in the previous menu.

Wait for the data reading to stabilize.

Click Next to complete the calibration.