Xylem WQ770-B TURBIDITY METER User Manual

Page 10

Advertising
background image

Global Water

800-876-1172

globalw.com

- 10 -

Calibration

a. If the sensor is not connected to the meter, push the cable connector

gently into the meter’s connector. The outer connector ring should be

turned and locked into place so the cable will not detach from the

meter during use.

b. To enter the meter’s calibration screens hold the Menu/On button for

3 seconds. The first menu screen is for selecting the range. Use the

Down arrow key to select 50 NTU then press the Save button.

c. Press the Menu/On button to display the units available. Use the

Up/Down arrow keys to select NTU (nephlometric turbidity units).

Press the Save button to record the unit selection. Press the Menu/On

button to move to the next menu.

d. The LoSig screen shows a number with NTU. This is not an indication

of the real NTU reading. This value includes meter offsets and other

variables.

e. You may use a lint free bent cotton swab to remove any air bubbles

from underneath the detector window, ensuring that there are no

bubbles on the lenses is critical. Place the sensor into the distilled

water. Press the Save button until the display reads “saved”. This

step may need to be repeated to ensure that an accurate reading is

taken. Press the Menu/On button to move to the next menu.

f. The HiSig screen shows a number with NTU. This is not an indication

of the real NTU reading. This value includes meter offsets and other

variables.

g. Place the sensor in the clamp above the 50 NTU solution. Lower the

clamp and dip the sensor into the solution. Wipe the detector window

with a cotton swab to remove any air bubbles. Press the Save button

until the display reads “saved”. This step may need to be repeated to

ensure that an accurate reading is taken. Press the Menu/On button

to move to the next menu.

h. The LoLab screen displays the correct NTU reading for the solution

used in the LoSig calibration step. For distilled water press the arrow

keys until the display reads 0 NTU. Press the Save button to record

the LoLab unit. Press the Menu/On button to move to the next menu.