III. WQ770 Operation

a. Before turning the unit on, make sure that the sensor cable has

correctly been attached to the display. Push the cable connector

gently into the meter’s connector. The outer connector ring can be

turned and locked into place so the cable will not detach from the

meter during use.

b. Remove the battery cover from the meter and verify that a 9V lithium

battery is installed in the unit. Replace the battery cover.

c. Press the Menu/On Button. This brings up the display that shows the

output from the turbidity sensor.

d. Press and hold the Menu/On button for 3 seconds to display the

Range menu. Use the Up/Down arrow buttons to select either the

0-50 or the 0-1000 NTU range then press Save.

e. Press the Menu/On button again and use the Up/Down arrow keys to

select units of NTU or ppm then press Save. Pressing the Menu/On

button 6 more times returns to normal display mode. The other menu

items are used for calibration and are described in Appendix A.

f. Place the sensor in the water to be tested. The meter will display the

turbidity of the water. Warning: Air bubbles in the water or under the

sensor shield will cause errors in the readings. These air bubbles

must be avoided when placing the sensor in the water.

g. The Turbidity Meter will automatically shut off after 6 minutes. To

continue reading the display past this time press the Menu/On button

prior to the meter shutting off.

h. Prolonged immersion in hot water will cause accelerated aging of the

sensor optics and seals. Warning: Do not expose the sensor to

water exceeding 50°C.

IV.

Battery Life And Replacement