Ix. appendix b: turbidity standard solutions, 003 0.0 ntu turbidity standard, 004 50 ntu turbidity standard – Xylem WQ770-B TURBIDITY METER User Manual

Page 11: 005 100 ntu turbidity standard, 006 1000 ntu turbidity standard

Advertising
background image

Global Water

800-876-1172

globalw.com

- 11 -

i. The HiLab screen displays the correct NTU reading for the solution

used in the HiSig calibration step. For 50 NTU solution, press the

arrow keys until the display reads 50 NTU. Press the Save button to

record the HiLab unit. Press the Menu/On button to move to the next

menu.

j. The delay screen indicates the time the meter will take to average

sensor readings. This should be set 5 seconds for the best results.

Press the Menu/On button to return to the initial measuring screen.

k. Repeat steps b-j to calibrate the 1000 NTU range using the 1000 NTU

solution and distilled water. Use the Up arrow button to select the

1000 NTU range in step b and set the HiLab value to 1000 in step i.

IX.

Appendix B: Turbidity Standard Solutions

The following turbidity standards are available from Global Water, contact the Sales

Department or Technical Support for more information.

These are EPA approved primary standards that are safe, non-toxic and disposable.

They are easy-to-use with no dilution or re-suspension needed. Their inherent

stability means they do not settle from suspension and require no stirring. The

convenient submicron polymer suspensions are non-toxic primary calibration

standards approved by the US EPA for drinking water analysis for turbidity, and

preferred by the U. S. Geological Survey for turbidity and sediment transport

analysis.

Part Number

Description

02-003

0.0 NTU Turbidity Standard

02-004

50 NTU Turbidity Standard

02-005

100 NTU Turbidity Standard

02-006

1000 NTU Turbidity Standard