Xylem WQ770-B TURBIDITY METER User Manual

Page 9

Advertising
background image

Global Water

800-876-1172

globalw.com

- 9 -

Preparation:

a. Set up a test stand with a rod and clamp to hold the sensor. Beakers

will be placed underneath the clamp so the sensor can be inserted into

the liquid.

b. The sensor must have the stainless steel section of the body in contact

with the test solution. Install the sensor so that the end of the sensor

is at least 3 inches from the bottom. If the sensor is too close to the

bottom, infrared light may be reflected from the glass and effect the

measurements.

c. The lenses must be gently cleaned with a cotton swab and isopropyl

alcohol before each calibration step.

d. Fill a large beaker with distilled water. This will be used to rinse

solution from the sensor in between steps.

e. Fill three more beakers with 0.0 NTU Standard, 50 NTU Standard and

1000 NTU Standard. These beakers must be at least 500 ml in size or

the end of the sensor will be too close to the bottom for accurate

calibration. This generally requires solution volumes of at least 500ml.

f. The calibration process should not be done in bright light if possible.

While not generally required, the extra step of covering the outside of

the beakers with a non-transparent material may help to prevent stray

light from the outside affecting the readings.

Note: The sensor is factory calibrated and will provide output currents

of 20mA at the maximum turbidity values; however, some small drift in

this output current can occur over time. These adjustments are done

internally, access to these settings are not available to the user. The

sensors should never be disassembled. The following process

describes how to re-calibrate the meter and sensor to compensate for

any drift.

Advertising