Ys a, Ol s – Xylem inoLab 7110 WASTEWATER pH METER User Manual

Page 23

Advertising
background image

pH 7110

pH value

ba75925e02

12/2011

23

3.

If necessary, press <CAL> to select the buffer set ([ConCal]).
ASY

is displayed.

4.

Thoroughly rinse the pH electrode with deionized water.

5.

Immerse the pH electrode in the first buffer solution (pH 7.0 ± 0.5 for
two-point calibration).

6.

When measuring without temperature sensor: Enter the temperature of
the buffer with <><>.

7.

Start the measurement with <ENTER> .
The pH value of the buffer solution is displayed.
The measured value is checked for stability (stability control).
The [AR] status display flashes.

8.

Wait for the measurement with stability control to be completed.

9.

Set the nominal pH value of the buffer solution with <><>.

10. Accept the calibration value with <ENTER>.

SLO

is displayed.

11. If necessary, finish the calibration procedure as a single-point calibration

with <M>.
The calibration record is displayed.

or
Continue calibration using the next buffer with <ENTER>.

Continuing with two-point calibration

12. Thoroughly rinse the pH electrode with deionized water.
13. Immerse the pH electrode in the second buffer solution.

8

4

2

Y

S

A

°C

TP

ConCal

8

4

2

O

L

s

°C

TP

ConCal

Advertising