3 error message ofl, ufl, 2 orp, 1 no stable measured value – Xylem inoLab 7110 WASTEWATER pH METER User Manual

Page 39: Error message ofl, ufl, No stable measured value

Advertising
background image

pH 7110

What to do if...

ba75925e02

12/2011

39

10.1.3

Error message OFL, UFL

The measured value is outside the measuring range.

The measured value is obviously incorrect.

10.2

ORP

10.2.1

No stable measured value

Cause

Remedy

• The used buffer solutions do not

agree with the set buffer set

• Set different buffer set

or

• Use different buffer solutions

• Buffer solutions too old

• Use only once.

Note the shelf life

• Buffer solutions depleted

• Change solutions

Cause

Remedy

• Cable broken

• Replace the electrode

• Gel electrolyte dried out

• Replace the electrode

• Air bubble in front of the junction

• Remove the air bubble

• Air in the junction

• Extract air or moisten the junction

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.