Multi 3430
What to do if...
105
ba75865e11
02/2011
6.2
pH
Error message,
OFL, UFL
Error message,
Error
Cause
Remedy
IDS pH sensor:
– Measured value outside the
measuring range
– Use suitable IDS pH sensor
– Air bubble in front of the junction
– Remove air bubble
– Air in the junction
– Extract air or moisten junc-
tion
– Cable broken
– Exchange IDS pH sensor
– Gel electrolyte dried out
– Exchange IDS pH sensor
Cause
Remedy
IDS pH sensor:
– The values determined for zero
point and slope of the IDS pH
sensor are outside the allowed
limits.
– Recalibrate
– Junction contaminated
– Clean junction
– IDS pH sensor broken
– Exchange IDS pH sensor
– Buffer solutions:
– The used buffer solutions are not
suitable for the selected buffer
set
– Select a different buffer set
or
– Use different buffer solu-
tions
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change solutions