Xylem Multi 3430 User Manual
Page 75

Multi 3430
Operation
75
ba75865e11
02/2011
Calibration evaluation of
the FDO
®
925
Calibration in
water vapor-
saturated air
Proceed as follows to calibrate the D.O. sensor:
Note
The sponge in the check and storage vessel must be moist (not wet).
Leave the sensor in the check and storage vessel long enough so it can
adapt to the ambient temperature.
Display
Calibration record
Relative slope
+++
S = 0.94 ... 1.06
++
S = 0.92 ... 0.94 or
S = 1.06 ... 1.08
+
S = 0.90 ... 0.92 or
S = 1.08 ... 1.10
Error
Error
S < 0.90 or
S > 1.10
Eliminate the error according to
chapter 6 W
1
Connect the D.O. sensor to the meter.
2
Place the FDO
®
925 D.O. sensor into the check and storage
vessel, FDO
®
Check.
3
Start the calibration with <CAL>.
The last calibration data (relative slope) is displayed.