Xylem Multi 3430 User Manual
Page 84

Operation
Multi 3430
84
ba75865e11
02/2011
Setting menu of
LR325/01
Menu item
Possible set-
ting
Description
Cell constant
0.090 ... ...
0.100 ...
0.110 cm
-1
The cell constant can be dis-
played and set here
Temp. comp. (TC) /
Method
nLF
lin
off
Procedure for temperature
compensation (see section
4.10.3).
This setting is only available
for the measured parame-
ters,
ϰ and
ρ
.
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear tem-
perature compensation.
This setting is only available
when the linear temperature
compensation is set.
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available
for the measured parame-
ters,
ϰ and
ρ
.
TDS factor
0.40 ... 1.00
Factor for TDS value
Reset
-
Resets all sensor settings to
the delivery condition (see
section 4.13.1)