Operation

Multi 3430

90

ba75865e11

02/2011

Settings

You can configure the automatic storage function with the following set-
tings:

Starting the automatic

storage function

To start the automatic storage function, select continue with <

S><T>

and confirm with <MENU/ENTER>. The meter switches to the mea-
sured value display.

ID number

1

Interval 30

s

Duration 180

min

Continue

0d03h00min

0

1d17h33min

Automatic data storage

22.09.2009 08:00

Set entire storing dura-
tion

Max. available storing
duration

Graphical display of the
storage usage

Menu item

Possible
setting

Description

ID number

1 ... 10000

ID number for the dataset series

Interval

1 s, 5 s, 10 s,
30 s, 1 min,
5 min,
10 min,
15 min,
30 min,
60 min

Storage interval.

The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.

Duration

1 min ... x
min

Storage duration.
Specifies after which time the auto-
matic storage should be terminated.

The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free storage locations.