VIII. Humidity Sensor

a. Humidity sensor specifications.

Output:

4-19mA

Range:

0-100% RH

Accuracy:

+ 2% RH

Operating Voltage:

10-36VDC

Current Draw:

3 mA plus sensor

Warm Up Time:

3 seconds minimum

Operating Temp:

-40° to +55°C

Size of Probe:

1 ½” diameter x 7” long

Weight:

½ lb.


b. The humidity sensor is a three-wire sensor. Three wire sensors use

the red wire for positive voltage, the white wire for the output signal,

and the black wire for ground. Warning: Always connect the

sensor with the power turned off.


c. Do not install the humidity sensor in direct sunlight.

d. A humidity sensor utilizes a thin polymer that varies in dielectric

constant directly proportional to changes in the amount of water vapor

at the sensor element. The element provides a linear voltage output

that is converted into a 4-19 mA sensor output signal.