Nadc/pdc (option e4406a-bae), Nadc/pdc, General characteristics – Atec Agilent-E4406A User Manual

Page 17

Advertising
background image

17

NADC/PDC

(Option E4406A-BAE)

ACPR measurement

The adjacent channel power ratio (ACPR) measurement
measures up to five pairs of offset channels and relates
them to the carrier power. The measurement result is a
ratio of the channel power to the power in each offset. The
results can be displayed as a ratio to the total power in
each bandwidth, or as a ratio of the power spectral density.

Carrier power range

+27 to –20 dBm

at RF input

Dynamic range

NADC mode

Offset frequency (Integ BW)

30 kHz (32.8 kHz)

–35 dB (nominal)

60 kHz (32.8 kHz)

–65 dB

90 kHz (32.8 kHz)

–70 dB

PDC mode

Offset frequency (Integ BW)

50 kHz (21.0 kHz)

–55 dB

100 kHz (21.0 kHz)

–70 dB

Relative accuracy

Resolution

±1.0 dB

Display resolution

0.01 dB

EVM measurement

EVM measurement measures the modulation quality of
pi/4QPSK modulated signal providing you with IQ constel-
lation diagram, error vector magnitude (EVM) in RMS and
peak as well as each chip of magnitude error, phase error
and EVM.

Range at RF input

+27 to –20 dBm

(Common in NADC
and PDC)

EVM

Range

0 to 25%

Floor

1.0%

Accuracy

±0.6%

I/Q origin offset

Range

–10 to –50 dBc

Resolution

0.01 dB display resolution

Carrier frequency error

Frequency resolution

0.01 Hz display resolution

OBW measurement (PDC only)

Occupied bandwidth (OBW) measurement measures the
frequency bandwidth corresponding to 99% of the total
transmitted power.

Range at RF input

+27 to –20 dBm

Frequency

Resolution

0.1 kHz

Accuracy

+400 Hz, –100 Hz

In-band frequency range (NADC)

800 MHz band

Mobile transmit

824 to 849 MHz

Base station transmit

869 to 894 MHz

PCS band

Mobile transmit

1850 to 1910 MHz

Base station transmit

1930 to 1990 MHz

In-band frequency range (PDC)

800 MHz band #1

810 to 828 MHz
940 to 958 MHz

800 MHz band #2

870 to 885 MHz
925 to 940 MHz

800 MHz band #3

838 to 840 MHz
893 to 895 MHz

1500 MHz band

1477 to 1501 MHz
1429 to 1453 MHz

General characteristics

Temperature range

Operating

0 °C to +55 °C

Non-operating

–40 °C to +71 °C

EMI compatibility

Conducted and radiated emission is in compliance with
CISPR Pub. 11/1990 Group 1 Class A.

Radiated immunity (RF input)

When tested at 3 V/m according to IEC 801-3/1984, the
displayed average noise level will be within specifications
over the full immunity test frequency range of 27 to
500 MHz, except that at immunity test frequencies
of 278.6 MHz ± selected resolution bandwidth and
321.4 MHz ± selected resolution bandwidth, the displayed
average noise level may be up to –90 dBm. When the
analyzer tuned frequency is identical to the immunity test
signal frequency there may be signals of up to ±90 dBm
displayed on the screen.

Advertising