Atec Agilent-E4406A User Manual

Page 9

Advertising
background image

9

Conformance with 3GPP TS 25.141 base station requirements for a manufacturing environment

Sub-clause

Name

3GPP required test

Instrument tolerance

Supplemental

instrument tolerance

interval

information

(as of June 2002)

6.2.1

Maximum output power

±0.7 dB (95%)

±0.29 dB (95%)

±0.63 dB (100%)

6.2.2

CPICH power accuracy

±0.8 dB (95%)

±0.30 dB (95%)

–10 dB CDP

6.3.4

Frequency error

±12 Hz (95%)

±10 Hz (100%)

Freq ref locked

6.4.2

Power control steps

1-dB step

±0.1 dB (95%)

±0.03 dB (95%)

Test Model 2

0.5-dB step

±0.1 dB (95%)

±0.03 dB (95%)

Test Model 2

Ten 1-dB steps

±0.1 dB (95%)

±0.03 dB (95%)

Test Model 2

Ten 0.5-dB steps

±0.1 dB (95%)

±0.03 dB (95%)

Test Model 2

6.4.3

Power dynamic range

±1.1 dB (95%)

±0.50 dB (95%)

6.4.4

Total power dynamic range

±0.3 dB (95%)

±0.015 dB (95%)

Ref –35 dBm
at mixer

6.5.1

Occupied bandwidth

±100 kHz (95%)

±38 kHz (95%)

10 averages

6.5.2.1

Spectrum emission mask

±1.5 dB (95%)

±0.59 dB (95%)

Absolute peak

6.5.2.2

ACLR

5 MHz offset

±0.8 dB (95%)

±0.34 dB (95%)

±0.93 dB (100%)

10 MHz offset

±0.8 dB (95%)

±0.40 dB (95%)

±0.82dB (100%)

6.7.1

EVM

±2.5% (95%)

±1.0% (95%)

Range 15 to 20%

6.7.2

Peak code domain error

±1.0 dB (95%)

±1.0 dB (nominal)

Conditions

25 to 35 °C

Derived tolerances

95th percentile

100% limit tested

Calibration uncertainties included