Frequency – Atec Agilent-E4406A User Manual

Page 2

Advertising
background image

2

Frequency

Frequency range

RF input

7 to 314 MHz and 329 MHz
to 4 GHz

Baseband IQ inputs

0 Hz to 5 MHz

Frequency spans

Baseband IQ inputs

5 Hz to 5 MHz
(Baseband I or Q inputs)

10 Hz to 10 MHz
(Composite I/Q)

Frequency setting resolution

1 Hz

Frequency reference

Accuracy

±[(time since last adjustment x
aging rate) + temperature
stability + calibration accuracy]

Initial calibration accuracy ±5 x 10

-8

Settability

±2 x 10

-9

Aging rate

During any 24 hrs

±5 x 10

-10

(nominal)

following 24-hr warm-up

Per year

±1 x 10

-7

(nominal)

Temperature stability

±5 x 10

-8

variation from

frequency at +25 °C over the
temperature range of 0 to +55 °C

Warm-up time

1 hour (nominal)

Residual responses

RF input

50

input terminated, 0 dB input attenuation,

+18 dB ADC gain

20 MHz to 2 GHz

–85 dBm

2 GHz to 4 GHz

–80 dBm

Baseband IQ inputs

50

input terminated

0 to 5 MHz

–90 dBm

Noise Sidebands (RF Input)

673.6 MHz

Offset

Specifications

Supplemental

100 Hz

–85 dBc/Hz

1 kHz

–92 dBc/Hz

10 kHz

–102 dBc/Hz

100 kHz

–131 dBc/Hz

600 kHz

–138 dBc/Hz

1.2 MHz

–141 dBc/Hz

6.0 MHz

–145 dBc/Hz

10.0 MHz

–145 dBc/Hz

960 MHz

Offset

Specifications

Supplemental

100 Hz

–81 dBc/Hz

1 kHz

–87 dBc/Hz

10 kHz

–96 dBc/Hz

100 kHz

–125 dBc/Hz

600 kHz

–136 dBc/Hz

1.2 MHz

–140 dBc/Hz

6.0 MHz

–146 dBc/Hz

10.0 MHz

–146 dBc/Hz

1990 MHz

Offset

Specifications

Supplemental

100 Hz

–75 dBc/Hz

1 kHz

–82 dBc/Hz

10 kHz

–86 dBc/Hz

100 kHz

–118 dBc/Hz

600 kHz

–132 dBc/Hz

1.2 MHz

–137 dBc/Hz

6.0 MHz

–141 dBc/Hz

10.0 MHz

–141 dBc/Hz

Noise Sidebands

1

(Baseband IQ Inputs)

0 to 5 MHz

Offset

Specifications

Supplemental

1 kHz

–120 dBc/Hz

10 kHz

–133 dBc/Hz

100 kHz

–134 dBc/Hz

1.0 MHz

–135 dBc/Hz (nominal)

5.0 MHz

–135 dBc/Hz (nominal)

1.

No DC offset applied

Advertising