W-cdma (option e4406a-baf), W-cdma – Atec Agilent-E4406A User Manual

Page 6

Advertising
background image

6

W-CDMA

(Option E4406A-BAF)

Channel power measurement

The channel power measurement measures the total RMS
power in a user-specified bandwidth. The following specifi-
cations apply for the default bandwidth of 3.84 MHz for the
3GPP standard.

Minimum power at

–70 dBm (nominal)

RF input

Absolute power accuracy, ±0.63 dB
18 to 30 °C

(±0.41 dB, typical)

Measurement floor

-73 dBm (nominal)

ACPR measurement (ACLR)

The adjacent channel power ratio (ACPR) measurement
measures up to five pairs of offset channels and relates
them to the carrier power. The measurement result is a
ratio of the channel power to the power in each offset.
The results can be displayed as a ratio to the total power
in each bandwidth, or as a ration of the power spectral
density. Simulated spectrum analyzer mode is for those
who are accustomed to spectrum analyzers.

Minimum power at

–27 dBm (nominal)

RF input

ACPR accuracy

RRC weighted,
3.84 MHz noise bandwidth

Radio

Offset frequency Specification

MS (UE) 5 MHz

±0.20 dB, at ACPR range of
–30 to –36 dBc with optimum
mixer level

MS (UE) 10 MHz

±0.30 dB, at ACPR range of
–40 to –46 dBc with optimum
mixer level

BTS

5 MHz

±0.93 dB, at ACPR range of
–42 to –48 dBc with optimum
mixer level

BTS

10 MHz

±0.82 dB, at ACPR range of
–47 to –53 dBc with optimum
mixer level

BTS

5 MHz

±0.39 dB, at –48 dBc
non-coherent ACPR

Dynamic range

RRC weighted,
3.84 MHz noise bandwidth

Offset frequency

5 MHz

–68 dB (nominal)

10 MHz

–72 dB (nominal)

For more detail, please refer to the E4406A specifications
that can be found at www.agilent.com/find/vsa

Power statistics CCDF measurement

The complementary-cumulative distribution function (CCDF)
traces provide you with how much time the waveform
spends at or above a given power level. The percent of time
the signal spends at or above the level defines the
probability for that particular power level.

Minimum power at

–40 dBm, average (nominal)

RF input

Histogram resolution

0.01 dB

Code domain measurement

The code domain measurement provides a tremendous
amount of information about the in-channel characteristics
of the W-CDMA signal. Code domain power (CDP) view
directly informs the user of the active channels with their
individual channel powers. The CDP view also leads you to
symbol rate analysis such as symbol rate EVM and symbol
power versus time.

Code domain power
25 to 35°C
95% confidence

Minimum power at

–70 dBm (nominal)

RF input

Relative code

Using Test Model 1 with

domain accuracy

32 DPCH signal

±0.015 dB

Code domain power between
0 and –10 dBc

±0.08 dB

Code domain power between
–10 and –30dBc

±0.15 dB

Code domain power between
–30 to–40dBc

Symbol power vs. time

Minimum power at

–45 dBm (nominal)

RF input

Accuracy

Using Test Model 1 with
32 DPCH signal

±0.10 dB

Code domain power
between 0 and –25 dBc

±0.50 dB

Code domain power
between –25 to –40dBc

Symbol error vector magnitude

Minimum power at

–45 dBm (nominal)

RF input

Accuracy

Using Test Model 1 with
32 DPCH signal

± 1.0%

Code domain power
between 0 and –25 dBc

Advertising