Compensation
For more details of the compensation functions, please see the instruction manuals supplied with the
measurement instruments or Application Note 346-3 (P/N 5091-6553E).
3. Electrical Length Compensation
In a single-port (Two-Terminal) impedance measurement at higher frequencies (RF region or higher),
the wavelengths of the electrical signal are so short that the length of the signal transmission line
including the test fixture and cables can cause an undesirable phase shift error. The phase shift
error is corrected by the electrical length compensation, which should be performed in combination
with the open/short compensation.
The following measurement instruments are equipped with electrical length compensation function.
The electrical length is specified for the following test fixtures:
Model
Compensation Function
4195A + 41951A
Test Port Extension
4286A
Electrical Length Compensation/Port Extension
4287A
Electrical Length Compensation/Port Extension
4291A/B
Electrical Length Compensation/Port Extension
4294A + 42942A
Electrical Length Compensation/Port Extension
4395A + Option 010 + 43961A
Electrical Length Compensation/Port Extension
4396A/B + Option 010 + 43961A
Electrical Length Compensation/Port Extension
E4991A
Electrical Length Compensation/Port Extension
Model
Electrical Length
16092A
3.4 mm
16093A
3.4 mm
16093B
3.4 mm
16094A
23.2 mm
16191A
14.0 mm
16192A
11.0 mm
16193A
14.0 mm
16194A
50.0 mm
16196A
26.2 mm
16196B
26.9 mm
16196C
27.1 mm
16197A
14.0 mm
Electrical length of a test fixture
Residual parameters in the measuring circuit