Appendix

Compensation

For more details of the compensation functions, please see the instruction manuals supplied with the
measurement instruments or Application Note 346-3 (P/N 5091-6553E).

3. Electrical Length Compensation

In a single-port (Two-Terminal) impedance measurement at higher frequencies (RF region or higher),
the wavelengths of the electrical signal are so short that the length of the signal transmission line
including the test fixture and cables can cause an undesirable phase shift error. The phase shift
error is corrected by the electrical length compensation, which should be performed in combination
with the open/short compensation.

The following measurement instruments are equipped with electrical length compensation function.

The electrical length is specified for the following test fixtures:

Model

Compensation Function

4195A + 41951A

Test Port Extension

4286A

Electrical Length Compensation/Port Extension

4287A

Electrical Length Compensation/Port Extension

4291A/B

Electrical Length Compensation/Port Extension

4294A + 42942A

Electrical Length Compensation/Port Extension

4395A + Option 010 + 43961A

Electrical Length Compensation/Port Extension

4396A/B + Option 010 + 43961A

Electrical Length Compensation/Port Extension

E4991A

Electrical Length Compensation/Port Extension

Model

Electrical Length

16092A

3.4 mm

16093A

3.4 mm

16093B

3.4 mm

16094A

23.2 mm

16191A

14.0 mm

16192A

11.0 mm

16193A

14.0 mm

16194A

50.0 mm

16196A

26.2 mm

16196B

26.9 mm

16196C

27.1 mm

16197A

14.0 mm

Electrical length of a test fixture

Residual parameters in the measuring circuit