Atec Agilent-16047A User Manual

Page 34

Advertising
background image

16314A Balanced/Unbalanced
4-Terminal Converter

Terminal Connector: 4-Terminal Pair, BNC

(unbalanced)

DUT Connection: 3 binding posts (balanced)

Dimensions (approx.):

89 (W) x 56 (H) x 133 (D) [mm]

Weight (approx.): 400 g

Additional Error:

(Typical Data)

29

Up to 110 MHz (4-Terminal Pair)

Balanced/Unbalanced Converters

Description: This balun converts the (unbalanced) 4-
Terminal pair configuration to a (balanced) binding posts
configuration. It is used to measure a balanced device
using an unbalanced measurement instrument.
Applicable Instruments: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, 4294A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: 100 Hz to 10 MHz
Nominal Characteristic Impedance:

Unbalanced Side: 50

Ω

Balanced Side: 50

Ω

Insertion Loss:

≤

1.0 dB (@ 23°C±5°C, 100 kHz)

Frequency Response:

≤

±1.0 dB (@ 23°C±5°C, 100 kHz)

specified @ 23°C±5°C

Operating Temperature: 0°C to 55°C
DUT Size: Terminal spacing
of balanced side: 14 mm

Furnished Accessories:

Compensation and Measurement: Open, short and
load compensations are recommended before measure-
ment. Short compensation is performed by shorting the
binding posts together with the furnished shorting plate.
Load compensation is performed by using the furnished
50

Ω

load resistor. After performing open, short and load

compensations, the DUT is connected to the binding posts
of the balun. If the DUT has a characteristic impedance
other than 50

Ω

, 16316A 100

Ω

Balanced/50

Ω

Unbalanced Converter or 16317A 600

Ω

Balanced/50

Ω

Unbalanced Converter can be used instead. To obtain
more details about these products, refer to their product
overview 5091-6748E.

Description

P/N

Qty.

50

Ω

load resistor

16315-60002

1

Shorting Plate

16315-60003

1

Operation and Service Manual

16315-90001

1

Frequency

Return Loss

100 Hz

≤

f < 300 Hz

≥

10 dB

300 Hz

≤

f

≤

7 MHz

≥

20 dB

7 MHz < f

≤

10 MHz

≥

17 dB

Frequency

Common Mode Loss

100 Hz

≤

f < 3 MHz

≥

50 dB

3 MHz

≤

f

≤

5 MHz

≥

45 dB

5 MHz < f

≤

10 MHz

≥

40 dB

4294A with 16314A

Frequency Range

Error

100 Hz

≤

f < 1 MHz

±0.2%

1 MHz

≤

f < 3 MHz

±0.5%

f

≥

3 MHz

±2%

Advertising