Up to 110 mhz (4-terminal pair) material – Atec Agilent-16047A User Manual
Page 44

Applicable Instruments: 4263B, 4268A, 4279A, 4284A, 4285A, 4288A, 4294A, (4192A, 4194A, 4263A,
4278A)*
* denotes the instrument is obsolete.
Frequency:
≤
30 MHz
Maximum Voltage: ±42 V peak max. (AC+DC)
Operating Temperature: 0°C to 55°C
Material Size:
39
Up to 110 MHz (4-Terminal Pair)
Material
Electrodes for contacting electrode method
(Rigid Metal Electrode)
Material size for electrode-A
Material size for electrode-B
Electrodes for contacting electrode method
(Thin Film Electrode)
Material size for electrode-C
Material size for electrode-D
Electrode Type
Diameter of MUT
Thickness of MUT
Diameter of Electrode
Max. Frequency
A
40 mm ~ 56 mm
t
≤
10 mm
38 mm
30 MHz
B
10 mm ~ 56 mm
t
≤
10 mm
5 mm
30 MHz
Electrode Type
Diameter of MUT
Thickness of MUT
Diameter of Electrode*
Max. Frequency
C
56 mm
t
≤
10 mm
5 ~ 50 mm
30 MHz
D
20 mm ~ 56 mm
t
≤
10 mm
5 ~ 14 mm
30 MHz
Equipped with Electrodes A and B for flat and smooth materials.
Equipped with Electrodes C and D for rough or extremely thin materials.
* diameter of applied thin film electrode