Atec Agilent-E8364B User Manual

Page 112

Advertising
background image

108

Table 35 (Continued). Test Port Input

Direct Receiver Access Input Noise Floor

a,b

10 Hz IF Bandwidth

10 MHz to

45 MHz

--

<-127 dBm

--

<-127 dBm

--

45 MHz to 500
MHz

c

--

<-127 dBm

--

<-127 dBm

--

500 MHz to

2 GHz

--

<-133 dBm

--

<-133 dBm

--

2 GHz to

10 GHz

--

<-132 dBm

--

<-132 dBm

--

10 GHz to

20 GHz

--

<-134 dBm

--

<-134 dBm

--

20 GHz to

40 GHz

--

<-125 dBm

--

<-125 dBm

40 GHz to

50 GHz

--

<-123 dBm

--

<-123 dBm

Option 016
degrades
performance by
2 dB.

1 KHz IF Bandwidth

10 MHz to

45 MHz

--

<-107 dBm

--

<-107 dBm

--

45 MHz to

500 MHz

c

--

<-107 dBm

--

<-107 dBm

--

500 MHz to

2 GHz

--

<-113 dBm

--

<-113 dBm

--

2 GHz to

10 GHz

--

<-112 dBm

--

<-112 dBm

--

10 GHz to

20 GHz

--

<-114 dBm

--

<-114 dBm

--

20 GHz to

40 GHz

--

<-105 dBm

--

<-105 dBm

40 GHz to

50 GHz

--

<-103 dBm

--

<-103 dBm

Option 016
degrades
performance by
2 dB.

Advertising