Atec Agilent-E8364B User Manual
Page 55
Advertising

51
Table 17. 85050C Calibration Kit
Standard Configuration and Standard Power Range
(E836xB/C)
Applies to the, E836xB/C analyzers, 85050C (7mm) calibration kit, 85132F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
0.045 to
2 GHz
2 to
18 GHz
Directivity
52
60
Source Match
48
60
Load Match
52
60
Reflection Tracking
±0.003
+0.02/°C
±0.000
+0.02/°C
Transmission Tracking
±0.004
+0.02/°C
±0.004
+0.02/°C
Advertising
This manual is related to the following products: