Table 23. 85054D Calibration Kit
Standard Configuration and Standard Power Range
(E836xB/C)
Applies to the, E836xB/C analyzers, 85054D (Type-N) calibration kit, 85132F flexible test port cable set with 85130C adapter set, and a full
2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
0.045 to
2 GHz
2 to
18 GHz
Directivity
40
34
Source Match
39
29
Load Match
40
34
Reflection Tracking
±0.003
+0.02/°C
±0.027
+0.02/°C
Transmission Tracking
±0.013
+0.02/°C
±0.136
+0.02/°C