Atec Agilent-E8364B User Manual
Page 49
Advertising

45
E836xB/C Corrected System Performance with 7mm Connectors
Table 15. 85050B Calibration Kit
Standard Configuration and Standard Power Range
(E836xB/C)
Applies to the, E836xB/C analyzers, 85050B (7mm)
calibration kit, 85132F flexible test port cable set, and a full 2-port
calibration. Also applies to the f
ollowing condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
0.045 to
2 GHz
2 to
18 GHz
Directivity
52
52
Source Match
48
41
Load Match
52
47
Reflection Tracking
±0.003
+0.02/°C
±0.047
+0.02/°C
Transmission Tracking
±0.004
+0.02/°C
±0.032
+0.02/°C