Loss/Length Testing in Smart Remote Mode (cont.)
6
Clean the tester's and source's OUTPUT connectors and
the reference test cord connectors.
7
On the main tester: press
F
; then select Set
Loss/Length Reference.
8
Connect the reference test cords as shown on the screen
and in Figure 25; then press
t
.
9
Optional: From the TEST SETUP screen, you can enter
reference test cord lengths to meet TSB-140 reporting
requirements:
Use
L
to highlight a reference test cord number, press
t
; then enter a length. Press
N
when you are done.
10
Press
C
OK to leave the TEST SETUP screen.
W
Caution
If you disconnect the reference test cords from the
tester's or remote's output, you must set the
reference again to ensure valid measurements.
-continued-