78
Using the Visual Fault Locator
1
Clean the connectors on the patch cord, if used, and the
fiber to be tested.
2
Connect the fiber directly to the tester’s VFL port or
connect using the patch cord.
3
Turn on the VFL as follows:
•
Press
D
VFL from the HOME screen.
or
•
Press
F
; then select Visual Fault Locator.
4
To toggle between continuous wave and pulse modes,
press
B
.
5
To turn the VFL off and on, press
D
.
6
Look for the red light to locate fibers or faults, as shown
in Figure 34.
Tip: To more quickly locate the VFL's light at a patch panel,
view the light indirectly by holding a white card or
paper in front of the fiber connector emitting the light.
Note
The locator’s light may not be visible through dark-
colored fiber jackets.