Example: The output current is set at 300 mA.

CURR 0.300<CR>

3. SAVE<CR>

The programmed values of the output current and voltage are saved. If the
power supply is restarted, the values of the output current and voltage will
be those previously saved.

If the Output On-Off button is pressed at power on, then the power supply
will start with Output Off state regardless of the previously saved state.

If the output current and voltage values are adjusted through the front
panel knobs or the state is modified by Output On-Off button, the values are
automatically saved after 2s from the last adjustment.

4. VOLT?<CR>

This command returns the voltage measured to the output terminals of the
power supply: xx.xxV<CR>

The voltage value returned is the same as the one displayed on the voltage
display.

If the power supply is in the Output Off state, then this command will return
the OFF<CR> message.

5. CURR?<CR>

This command returns the current measured to the output terminals of the
power supply: x.xxxA<CR> or xx.xxA<CR>

The current value returned is the same as the one displayed on the current
display.

If the power supply is in the Output Off state, then this command will return
the OFF<CR> message.