B&K Precision 1739 - Manual User Manual

Page 28

Advertising
background image

26

b) Connect a digital ammeter (BK Precision 5491B or similar instrument) to the

output terminals of the power supply to initiate the full current calibration
procedure. If a digital ammeter is not connected within 30 seconds, the
power supply will abandon the current calibration procedure and will return
to the Calibration menu.

The CV LED indicator will turn off and the CC LED indicator will light on. The
power supply will then display:


Voltage display: I F

Current display: hex number

A hexadecimal number between 0 and FFFFh is shown on the current
display.

c) Use the current coarse and fine knobs to adjust the value displayed by the

power supply until the ammeter indicates the value to 1015.8mA.

d) Press the Output On-Off button to finish the calibration.

While performing the full current calibration, the current display will show
Adc. After the full current calibration, the message DonE is shown on the
current display for 1 sec. The power supply does not return to the
Calibration menu because the calibration procedure is finished and the
power supply must be turned off.

6. Leaving the calibration menu

a) After turning off the power supply, disconnect the power cord.

b) When turned on, the power supply will have the same parameters

programmed and saved before the calibration procedure.