7.3 Command Errors

Communication Error<CR>: RS232 framing, parity, or overrun error
“Syntax Error<CR>” : invalid syntax was found in the command string
“Out of range<CR>” : a numeric parameter value is outside the valid range

for the command

8 Specifications

1739

Output Ratings ( 0 °C~40 °C)
Voltage

0-30 V

Current

0-999.9 mA

Load Regulation ±(% of output+offset)
Voltage

0.04% + 1 mV

Current

0.4% + 0.1 mA

Line Regulation ±(% of output+offset)
Voltage

0. 1% + 3 mV

Current

0.4% + 0.3 mA

Ripple & Noise (20 Hz ~ 20 MHz)
Voltage

<1 mVrms

Current

≤ 0.4 mArms

Recovery Time

≤ 75

μs

Metering Range
Voltage

0-99.99 V

Current

0-999.9 mA

Metering Accuracy
Voltage

0.5% + 2 digits

Current

0.5% + 5 digits

General
Power Requirements

110/220 VAC ±10%, 50/60 Hz

Power Consumption

≤ 70 W

Protection

Reverse polarity, current limiting

Operating Environment